Emerson Process Management MON2000 Whiteboard Accessories User Manual


 
5-60 APPLICATION FUNCTIONS
MON2000
Valves JULY 2010
See the following table for details.
3. Click the button to accept your
changes and return to the main window.
Click the button to abort and
return to the main window.
Table 5-12 Description of Valve Settings
Setting Description
Channel channel number assigned
Label name used to identify the valve (12 characters maximum)
Switch sets operational mode for this valve
Auto = valve operation is determined by the GC application
Off = valve is set to OFF
On = valve is set to ON
Usage
(Model 700 and
Model 500 GCs with
2350A Controllers)
sets which digital output from the Timed Events function
(see Section 5.3) is assigned to this valve
Each Usage option is created per the following example:
1-AV1 - where the digit “1” is taken from the valve number set by the
Valve/DO# option (see Timed Events, see Section 5.3), and the name
“AV1” is taken from Label set in this Valve dialog.
Current Value current state of the valve
By default, the 2350A GC stream valves are shown first and are
labeled S1 - S12 followed by the analytical valves (AV1 - AV5). Model
700 GCs analytical valves are shown first and are labeled AV1 - AV5
followed by the stream valves (S1 - S8).
The same Usage assignment cannot be used for multiple valves,
unless these valves are set to “Unused”.