HP (Hewlett-Packard) HP 83236B PCS Interface Network Router User Manual


 
29
Chapter 2, Making Measurements
Modulation Quality (Rho)
Steps 1 Connect the equipment as shown in figure 8.
2 In the main menu, enter the Base Station Channel number for the transmitter
being tested.
3 Also in the main menu, enter the Base Station PN Offset.
4 Set up the base station to transmit (pilot only) on the channel number entered in step 2.
5 Verify that the REF UNLOCK light on the front of the Cellular Adapter is not lit.
6 Scroll to and select Measure Rho Parameters screen (see figure 9).
7 Read the value under the Rho field.
8 Read the values after the Time offset, Freq error, and Carrier ft fields.
9 Press the Main Menu (k5) key to return to the main menu.
Figure 9
Notes If the message CANNOT CORRELATE appears on the screen, see "CDMA
Measurement not Correlating" in chapter 4, "Troubleshooting, on page 69."
For an on-screen summary of the possible causes and solutions, press the
Diagnose (k4) key.