HP (Hewlett-Packard) HP 83236B PCS Interface Network Router User Manual


 
66
Chapter 4, Troubleshooting
Errors When Loading and Running the Software
Test System
Configuration
Errors at Runtime
Problem:
When it is first run, the Software checks to see that all required elements of a
CDMA/PCS Test System are present. If it does not find a CDMA Cellular
Adapter or a PCS Interface, it will report the following errors:
HP-IB Command not accepted. Option not installed.:
This error will be reported if the Software cannot “find” the CDMA Cellular
Adapter as an installed option on the Test Set.
Solution:
1 Make sure that your are using either the HP 83203B or HP 83205A CDMA Cellular
Adapter in your Test System.
2 Verify that you have supplied the Cellular Adapter with AC power. The front-panel
PWR light should be illuminated.
3 If the unit is powered on and you still get this message, you will need to check the rear-
panel connections between the Test Set and Cellular Adapter, especially the
CONTROL I/O cable between the two units. Refer to the Installation chapter of the
Cellular Adapter’s manual for connections.
Verify PCS Interface is connected and on. Retry?:
This error will be reported if the Software cannot “find” the HP 83236B PCS
Interface on the HP-IB bus.
Solution:
1 Verify that you have supplied the PCS Interface with AC power and that the POWER
switch is on. The front-panel POWER light should be illuminated.
2 If the unit is powered on and you still get this message, check the rear-panel HP-IB
connection between the Test Set and Cellular Adapter.
3 Confirm that the rear-panel HP-IB Address Selector switch is set to “HP-IB” (not
“Ser”). If the switch is in the wrong position, correct it and cycle power on the PCS
Interface.