BioOptix™ 10
Section 2 Installation
2-7
10. Connect the tubing from the column module to the pump
module. Refer to Figure 2-10.
a. Remove the two thumbscrews securing the splash
guard to the pump module. Set aside the splash guard.
Note
The tubing numbers from the column module match the num-
bers above the compression fittings on the pump module. (The
tubing is visible in Figure 2-9; the compression fittings are visi-
ble in Figure 2-10.)
b. Loosen the nut on the compression fitting, insert the
tubing fully into the fitting and finger-tighten the nut.
Then, use a wrench to tighten the nut an additional full
turn. Repeat for all ten channels.
c. Replace the splash guard so that the tubing is routed
up and over the guard.
Figure 2-10 Liquid connections
(pump module shown)
11. Slide the two modules together (Refer to Figure 2-11).
Align the eight holes on the upper and side case panels,
and secure with screws.
Note
Use care to avoid damaging internal tubing and cables. Do not
pinch or crimp tubing or cables between modules.
Thumbscrew
Splash Guard
Fittings for column
module tubing