BioOptix™ 10
Section 2 Installation
2-11
Figure 2-14 Buffer assembly routes a common solution to
all ten channels
Figure 2-15 Buffer assemblies installed
To connect the Buffer Assembly:
a. Remove the two screws located in the center of the
Buffer A inlet ports. (The two exposed holes are used to
mount the Buffer Assembly.)
b. Attach the mounting block to the two holes using the
hex head screws.
c. Attach the manifold to the mounting block with the
Phillips screws and lock washers.
d. Connect the ten Teflon tubes (each 3" long by 0.085 ID)
from the manifold to the Buffer A inlet ports.
e. Cut to length a section of 0.250" OD Teflon tubing (P/N
023-0508-01) from the accessory kit and route the
buffer solution to the common inlet. The standard
length is 60" or 152 cm.
Common Buffer Inlet
Manifold
Mounting
Block