Remote Operation
4–24 TM-XDOP-01XN
Power On Output State
You may also change the output state whether the output is enabled or disabled at
power on.
To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]
To check the setting:
OUTP:PON:STAT?
Reset
Resetting the unit puts certain features to a known state. These states are listed in
the table below.
To reset the unit:
*RST or SYST:RES
Table 4-8
Features Affected by Reset (*RST) Command
Feature Reset State (*RST)
Voltage setpoint 0.0V
Current setpoint 0.0A
Power setpoint 103% of power rating
Low voltage setpoint limit 0.0V
High voltage setpoint limit 103% of voltage rating
Low current setpoint limit 0.0A
High current setpoint limit 103% of current rating
Low power setpoint 0% of power rating
High power setpoint 103% of power rating
Over voltage protection Disabled (0.0V)
Under voltage protection Disabled (0.0V) and not shutdown when tripped
Over current protection Disabled (0.0A) and not shutdown when tripped
Under current protection Disabled (0.0A) and not shutdown when tripped
Over power protection Disabled (0.0W) and not shutdown when tripped
Under power protection Disabled (0.0W) and not shutdown when tripped
Fold shutdown protection None and delay 0.5s
AC off shutdown recovery Autorecover
Over temperature shutdown
recovery
Latched
Triggered voltage setpoint Disabled (Default)
Triggered current setpoint Disabled (Default)
TM-XDOP-01XN.book Page 24 Monday, July 17, 2006 11:19 AM