Remote Operation
4–34 TM-XDOP-01XN
Configure Auxiliary Status Lines
To set up Auxiliary line A:
OUTP:AUXA:SOUR <aux-line-mnemonic>
The choices for <aux-line-mnemonic> include:
• NONE
• ON: Output On
• OFF: Output Off
• OVOL: Over-Voltage Condition
• UVOL: Under-Voltage Condition
• OCUR: Over-Current Condition
• UCUR: Under-Current Condition
• OPOW: Over-Power Condition
• UPOW: Under-Power Condition
• ACOF: AC Power had been turned off or failed
• OTEM: Over-Temperature Condition
• HTEM: High-Temperature Condition
• SPR: Sense Protection Tripped
• FOLD: Fold Protection Tripped
• CC: Constant Current Mode
• CV: Constant Voltage Mode
• CP: Constant Power Mode
• UNR: Output Unregulated
To check the setting:
OUTP:AUXA:SOUR?
To set up Auxiliary line B:
OUTP:AUXB:SOUR <aux-line-mnemonic>
To set the polarity of the auxiliary status lines:
OUTP:AUXA:POL {HIGH|LOW}
OUTP:AUXB:POL {HIGH|LOW}
Where:
HIGH means that the logic of the output is active high. (That is, if the condition is
true, the line is pulled high.)
LOW means the logic of the output is active low.