SCPI Command Reference
C–16 TM-XDOP-01XN
Table C-5
Commands for Calibration
Function SCPI Command Description Query
Restore Factory
Calibration
[:]CALibration[<channel>]:RESTore
Restore the calibration to the
constants set at the factory
N/A
Change Calibration
Password
[:]CALibration[<channel>][:SECure]:CODE
<codeword>
Change the calibration security
code.
No
Set Calibration
State
[:]CALibration[<channel>][:SECure]:STATe <on-
off-state>,<codeword>
Change calibration state (mode) Yes
Set Analog Current
Programming Input
Level
[:]CALibration[<channel>]:ANALog:<5V|10V>:PROG
ram:CURRent:LEVel {MINimum|MAXimum}
Set analog programming current
calibration level
No
Enter Analog
Current
Programming Input
Data
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:CURRent[:DATA] <current>
Set analog programming current
calibration data
No
Set Analog Voltage
Programming Input
Level
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:VOLTage:LEVel
{MINimum|MAXimum}
Set analog programming voltage
calibration level
No
Enter Analog
Voltage
Programming Input
Data
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:PROGram:VOLTage[:DATA] <current>
Set analog programming voltage
calibration data
No
Set Analog Current
Readback Output
Level
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:READback:CURRent:LEVel
{MINimum|MAXimum}
Set analog readback current
calibration level
No
Enter Analog
Current Readback
Output Data
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:READback:CURRent[:DATA] <current>
Set analog readback current
calibration data
No
Set Analog Voltage
Readback Output
Level
[:]CALibration[<channel>]:ANALog:
:{5V|10V}:READback:VOLTage:LEVel
{MINimum|MAXimum}
Set analog readback voltage
calibration level
No
TM-XDOP-01XN.book Page 16 Monday, July 17, 2006 11:19 AM