SCPI Commands for Digital Interfaces
TM-XDOP-01XN 4–33
AC Off Protection
The user has the option of setting whether the AC Off protection mechanism is
latched or automatically resumes operation.
SENSE:VOLT:AC:PROT:LATCH {ON|1|OFF|0}
Where:
ON or 1 means the supply will be latched in shut down if an AC Off condition
occurs, until the user re-enables the output.
and
OFF or 0 means the supply will automatically resume operation when the AC line
input returns to normal.
The default value is OFF.
To query if AC Off has tripped:
SENSE:VOLT:AC:PROT:TRIP?
Clear Protection Event
To clear a protection mechanism that has tripped:
OUTP:PROT:CLE
This will clear all protection mechanisms and re-enable the output. If the
condition that caused the alarm still exists, the protection will be allowed to trip
again.
View Power Supply Output
The following 3 commands query the voltage, current, or power being supplied at
the output terminals.
To measure the voltage output:
MEAS:VOLT?
To measure the current output:
MEAS:CURR?
To measure the power output:
MEAS:POW?
TM-XDOP-01XN.book Page 33 Monday, July 17, 2006 11:19 AM