Agilent Technologies E5810A Network Card User Manual


 
Chapter 4 115
Troubleshooting Information
Checking the E5810
Checking Front
Panel Display Error
Messages
This table shows E5810 front panel display error messages and typical
cause(s) for the error message.
Error Message Description
Firmware Update
not successful
The firmware update failed. Verify that the IP address of the ftp server,
login, user password and filename are correct.
ERROR: NV RAM
Setting defaults
Configuration nonvolatile data memory was corrupt. All data were
restored to factory defaults, per the nonvolatile RAM (Random Access
Memory).
ERROR: Selftest
See syslog
A power-on self-test error occurred. Errors may be viewed in the syslog
message by typing http://<IP Address>/systemLog.htm.
ERROR: No LAN
connection found
LAN Link integrity not detected. Make sure the LAN cable is connected
and the LAN is active.
ERROR: IP in use
<addr>
The E5810 tried to use an IP address that is currently being used by
another device on the network. Try these actions:
Reboot the E5810 with the Preset button depressed to set the
E5810 back to booting with DHCP enabled.
OR
Remove other device(s) from the network and reboot the E5810
OR
Move the E5810 to a Local network, disable DHCP, and set the
E5810 to an unused IP Address.