Agilent Technologies E5810A Network Card User Manual


 
116 Chapter 4
Troubleshooting Information
Checking the E5810
Checking E5810 Parameters
If the front panel display and LEDs indicate no error conditions, check the
E5810 configuration parameters.
Checking E5810
Current Parameters
If the E5810 is connected to an Enterprise (corporate) network, you can
check the E5810 current settings displayed on the
Configuring your E5810
LAN/GPIB Gateway
page against the Enterprise Network Information
provided to you by your System Administrator. If the settings do not match,
re-configure the E5810 from the
Configuring your E5810 LAN/GPIB Gateway
page.
Resetting E5810
Defaults
As required, you can use the button on the E5810 front panel to reset
the E5810 to its default settings. There are two modes of operation. See
Chapter 1 - E5810 Description for the location of the
Preset button.
Momentarily Depress the Preset Button. If you momentarily
depress the
button, ONLY the password is temporarily
set to the default value of E5810. All other E5810 configuration
parameters remain the same. The password remains until the
next time you reboot the E5810.
Depress the Preset Button for at Least 10 Seconds. If you
depress the
Preset button for at least 10 seconds, ALL E5810
configuration parameters are set to their defaults and the
E5810 reboots.
NOTE
If the E5810 is connected to an Enterprise (corporate) Network, be sure
to contact your IT Department before making any configuration changes
to the E5810. If E5810 configuration checks are required, see
Chapter 3 - Using E5810 Web Access for information.
NOTE
You should terminate all applications that use the E5810 before taking
this action.