Agilent Technologies E5810A Network Card User Manual


 
Chapter 4 127
Troubleshooting Information
Checking PC Client Connections
Example: Typical
syslog Messages
This table shows typical syslog messages and provides a description of
each message, where the value in ( ), such as (0:00:00:18), is the time since
power-on expressed in Day (DD):Hour (HH):Minutes (MM):Seconds (SS).
ERROR: Firmware Update:
FTP server not found
Could not find FTP server specified.
ERROR: Firmware Update:
Invalid User Login or Password
Login or Password is invalid for the FTP server.
ERROR: Firmware Update:
Data not received from FTP
Server
Could not receive data from the FTP Server. May be trying to
access FTP server through a firewall.
ERROR: Firmware Update:
Invalid file or directory
Could not find file. Filename or directory is wrong.
ERROR: Firmware Update:
Internal error
Out of system resources. Reboot the E5810 and try again.
Message Description
Message Description
(0:00:00:00) Selftest passed Power-on self-test results
(0:00:00:00) GPIB: Initialized, symbolic name = gpib0
GPIB and RS-232 initialized
(0:00:00:00) RS232: Initialized, symbolic name = COM1
(0:00:00:18) client: 156.140.104.184, id: 1 opened
Typical open/close messages (logged
each time a client opens or closes a
connection)
(0:00:00:18) client: 156.140.104.184, id: 1 closed
(0:00:00:18) client: 156.140.104.184, id: 1 opened
(0:00:00:19) client: 156.140.104.184, id: 2 opened
(0:00:00:19) client: 156.140.104.184, id: 3 opened
(0:00:39:21) Max connections exceeded, client refused Indicates the server had the maximum
number of connections when another
connection was requested, so the request