Agilent Technologies E5810A Network Card User Manual


 
Chapter 1 21
E5810 Description
E5810 Hardware Description
Power-On (Default)
Settings
When power is first applied to the E5810 and the hardware self-test has
completed, the E5810 is initialized to the factory-set default configuration.
This table shows the default configuration parameter settings for the E5810.
The E5810 uses these default configuration values until you set any other
configuration values. The E5810 also uses these values when you press the
Preset button and hold it down for >10 seconds. See Chapter 3 - Using
E5810 Web Access for a description of each parameter.
Preset Button
Operation
The Preset button on the front panel of the E5810 is used to reset the E5810
to its default configuration values (preset at the factory). As shown in the
figure on the next page, the
Preset button has two modes of operation:
If you depress the Preset button and release it in <10 seconds, only
the E5810 password is temporarily changed and is reset to its
default value (E5810). If you cycle power or reboot the E5810, the
E5810 will return to the actual password that was previously set. All
existing configuration values remain unchanged.
If you depress and hold the Preset button for 10 seconds or more, all
configuration values are reset to their factory default values and the
E5810 is rebooted. All network connections to the E5810, if any, are
also terminated without any cleanups.
Parameter Default Parameter Default
DHCP: ON GPIB Address: 21
IP Address: 169.254.58.10 GPIB Logical Unit: 7
Subnet Mask: 255.255.0.0 RS-232 SICL Interface Name: COM1
Default Gateway: 0.0.0.0 RS-232 Baud Rate: 9600
DNS Server(s): 0.0.0.0 RS-232 Parity: NONE
Hostname: none RS-232 Bits: 8
Description: Agilent E5810
<MAC Addr>
RS-232 Stop Bits: 1
Universal Plug & Play: ON RS-232 Flow Control: NONE
LAN Keepalive (sec): 7200 sec RS-232 SRQ: RI
I/O Timeout (sec): 120 sec Password E5810
GPIB SICL Interface Name: gpib0