5-12 Issue 05, 02/00 Krautkramer USN 52R/USN 52L
Calibration functions
USN 52R/USN 52L are accurately calibrated using the
functions of the function group S-CAL.
– With I go the function group S-CAL.
H Note:
Read Chapter 5.3 first in order to exactly set the sound
velocity and the probe delay.
H Note:
Refer also to the AUTO CAL function in Chapter 5.3.
The function group S-CAL correspondingly changes
when this function is used.
a-START Gate start
The active value for the gate start, which you have
entered into the function group GATE, is displayed in
a-START. If necessary, you can make a subsequent
change to this value here.
H Note:
In order to obtain the correct setting, you must have
selected the display range, according to the test task,
in the function group BASIC beforehand.
Operation Adjustment of the USN 52R/USN 52L
5-12 Issue 05, 02/00 Krautkramer USN 52R/USN 52L
Calibration functions
USN 52R/USN 52L are accurately calibrated using the
functions of the function group S-CAL.
– With I go the function group S-CAL.
H Note:
Read Chapter 5.3 first in order to exactly set the sound
velocity and the probe delay.
H Note:
Refer also to the AUTO CAL function in Chapter 5.3.
The function group S-CAL correspondingly changes
when this function is used.
a-START Gate start
The active value for the gate start, which you have
entered into the function group GATE, is displayed in
a-START. If necessary, you can make a subsequent
change to this value here.
H Note:
In order to obtain the correct setting, you must have
selected the display range, according to the test task,
in the function group BASIC beforehand.
Operation Adjustment of the USN 52R/USN 52L