Epson USN 52L Printer User Manual


 
5-22 Issue 05, 02/00 Krautkramer USN 52R/USN 52L
Higher sound velocity
Due to the V-path error, a higher sound velocity than
the test material is required when calibrating, especially
with smaller thicknesses. This is typical of dual (T/R)
probes and acts as compensation of the V-path error.
Owing to the described effect there is a drop in the
echo amplitude with smaller thicknesses which should
be carefully observed with values < 2 mm.
A calibration block having different thicknesses must be
used for calibration. The thicknesses are to be selected
so that they include the expected measurement value.
Set PULSER in the function group PULS to DUAL.
Set RANGE according to the test task and the probe
being used..
Set TOF to FLANK.
Couple the probe to the thinner part of the calibration
block and peak the echo.
Set the flank as steeply as possible by adjusting the gain.
Set Gate a on the first echo.
Set ZERO in the function group S-CAL so that the
measurement line display the correct value at
S=....
Couple the probe to the thicker part of the calibration
block and peak the echo.
Set the flank as steeply as possible by adjusting the
gain.
Set Gate a on the first echo.
Set the sound velocity C so that the correct value is
displayed in the measurement line at S=....
Automatic calibration
The instrument has an automatic calibration capability
via the function AUTO CAL. Two calibration echoes
must be entered. The instrument then carries out a plau-
sibility check and automatically sets the parameter.
Set the function AUTO CAL in the second operation
level (right function table) to ON.
Return to the first operation level with X and select
function group S-CAL with I.
The function group S-CAL changes in preparation for
the automatic calibration and now contains the CAL
function (calibration) as well as the function S-REF
(sound path for the reference echo).
Operation USN 52R/USN 52L calibration
5-22 Issue 05, 02/00 Krautkramer USN 52R/USN 52L
Higher sound velocity
Due to the V-path error, a higher sound velocity than
the test material is required when calibrating, especially
with smaller thicknesses. This is typical of dual (T/R)
probes and acts as compensation of the V-path error.
Owing to the described effect there is a drop in the
echo amplitude with smaller thicknesses which should
be carefully observed with values < 2 mm.
A calibration block having different thicknesses must be
used for calibration. The thicknesses are to be selected
so that they include the expected measurement value.
Set PULSER in the function group PULS to DUAL.
Set RANGE according to the test task and the probe
being used..
Set TOF to FLANK.
Couple the probe to the thinner part of the calibration
block and peak the echo.
Set the flank as steeply as possible by adjusting the gain.
Set Gate a on the first echo.
Set ZERO in the function group S-CAL so that the
measurement line display the correct value at
S=....
Couple the probe to the thicker part of the calibration
block and peak the echo.
Set the flank as steeply as possible by adjusting the
gain.
Set Gate a on the first echo.
Set the sound velocity C so that the correct value is
displayed in the measurement line at S=....
Automatic calibration
The instrument has an automatic calibration capability
via the function AUTO CAL. Two calibration echoes
must be entered. The instrument then carries out a plau-
sibility check and automatically sets the parameter.
Set the function AUTO CAL in the second operation
level (right function table) to ON.
Return to the first operation level with X and select
function group S-CAL with I.
The function group S-CAL changes in preparation for
the automatic calibration and now contains the CAL
function (calibration) as well as the function S-REF
(sound path for the reference echo).
Operation USN 52R/USN 52L calibration