Krautkramer USN 52R/USN 52L Issue 05, 02/00 5-31
5.6 Echo evaluation with
DAC/TCG
To be able to evaluate a flaw indication by means of the
DAC/TCG, certain requirements have to be met:
The Distance-Amplitude Curve must already be
recorded.
It
only
applies to the same probe that was used for
recording the curve. Not even another probe of the
same type must be used!
The curve and the DAC/TCG
only
apply to the
material that corresponds with the material of the
reference block.
All functions affecting the echo amplitude must be
set as they were during recording of the curve. This
especially applies to the following functions:
DAMPING, FREQ.,RECTIF., MTL VEL, PULSER
and REJECT.
TOF is set to PEAK.
AMPLITUDE is set to dB THRESHOLD.
Choose the echo amplitude (as dB difference to the
curve) using the T V -keys for an enlarged
representation in the A-scan.
Echo evaluation with DAC/TCG Operation
Krautkramer USN 52R/USN 52L Issue 05, 02/00 5-31
5.6 Echo evaluation with
DAC/TCG
To be able to evaluate a flaw indication by means of the
DAC/TCG, certain requirements have to be met:
The Distance-Amplitude Curve must already be
recorded.
It
only
applies to the same probe that was used for
recording the curve. Not even another probe of the
same type must be used!
The curve and the DAC/TCG
only
apply to the
material that corresponds with the material of the
reference block.
All functions affecting the echo amplitude must be
set as they were during recording of the curve. This
especially applies to the following functions:
DAMPING, FREQ.,RECTIF., MTL VEL, PULSER
and REJECT.
TOF is set to PEAK.
AMPLITUDE is set to dB THRESHOLD.
Choose the echo amplitude (as dB difference to the
curve) using the T V -keys for an enlarged
representation in the A-scan.
Echo evaluation with DAC/TCG Operation