Performance Tests
When received, the 2620A/2625A Hydra Series II instrument is calibrated and in
operating condition. The following Performance Verification Procedures are provided
for acceptance testing upon initial receipt or to verify correct operation at any time. All
tests may be performed in sequence to verify overall operation, or the tests may be run
independently.
If the instrument fails any of these performance tests, calibration adjustment and/or
repair is needed. To perform these tests, you will need a Fluke 5700A Multifunction
Calibrator or equipment meeting the minimum specifications given in Table 6-2.