Fluke 2620A Power Supply User Manual


 
2620A, 2625A
Users Manual
6-8
Table 6-3. Performance Tests (Voltage, Resistance, and Frequency)
Function Range Input Level Frequency
Display Accuracy
(1 year, 18-28°C)
MIN MAX
DC Volts
300 mV 0V
--
-0.02 0.02
300 mV 150 mV
---
149.94 150.06
300 mV 290 mV
---
289.91 290.09
3V 2.9V
---
2.8991 2.9009
-3V -2.9V
---
-2.9009 -2.8991
30V 29V
---
28.991 29.009
150V 150V
---
149.94 150.06
300V 290V
---
289.91 290.09
Note
Voltages greater than 150V can only be applied to channels 0,1, and 11.
300 mV
20 mV 1 kHz
19.71 20.29
300 mV
20 mV 100 kHz
18.50 21.50
300 mV
290 mV 1 kHz
289.28 290.72
3V
290 mV 100 kHz
275.00 305.00
3V
2.9V 1 kHz
2.8937 2.9063
30V
29V 1 kHz
28.931 29.069
150V
150V 1 kHz
149.54 150.46
300V
290V 1 kHz
289.34 290.66
Note
Voltages greater than 150V can only be applied to channels 0, 1, and 11. The rear Input Module must be
installed when measuring ac volts on channel 0.
Note
For 2-terminal measurements, the resistance accuracy given in this table applies to Channel 0 and makes
allowance for up to 0.05 ohm of lead wire resistance. You must add any additional lead wire resistance
present in your set up to the resistance values given in this table.
Resistance Using inputs in decades of 3:
300 short 0.00 0.09
300 299.90 300.15
3 k short 0.0000 0.0003
3 k 2.9989 3.0011
30 k 30 k 29.990 30.010
300 k 300 k 299.88 300.12
3 M 3 M 2.9979 3.0021