Fluke 2620A Power Supply User Manual


 
Appendices
Specifications
A
A-9
Ohms Inputs
Range Resolution
Typical Full
Maximum
Current
Through
Maximum Open
Slow Fast
Scale Voltage Unknown Circuit Voltage
300 10 m 0.1 0.22 V 1 mA 3.2V
3 k 0.1 1 0.25V 110 µA 1.5 V
30 k 1 10 0.29 V 13 µA 1.5 V
300 k 10 100 0.68 V 3.2 .µA 3.2 V
3 M 100 1 k 2.25 V 3.2 µA 3.2 V
10 M 1 k 10 k 2.72 V 3.2 µA 3.2 V
4-Wire Accuracy ±(%±)
Range 18°C to 28°C 0°C to 60°C
90 Days, Slow 1 Year, Slow 1 Year, Fast 1 Year, Slow 1 Year, Fast
300 0.013% + 20 m 0.014% + 20 m 0.014% + 0.2 0.031% + 20 m 0.031% + 0.2
3 k 0.015% + 0.2 0.016% + 0.2 0.016% + 2 0.039% + 0.2 0.039% + 2
30 k 0.013% + 2 0.014% + 2 0.014% + 20 0.039% + 2 0.039% + 20
300 k 0.020% + 20 0.021% + 20 0.021% + 200 0.050% + 20 0.050% + 200
3 M 0.059% + 200 0.063% + 200 0.063% + 2 k 0.231% + 200 0.231% + 2 k
10 M 0.168% + 2 k 0.169% + 2 k 0.709% + 20 k 0.573% + 2 k 0.923% + 20 k
2-Wire Accuracy
Add 4.0 to accuracy specifications for channels 1 to 20, and add 20 m for channel 0.
Lead wire resistances are not included.
Input Protection
300V dc or ac rms on all ranges
Cross-Talk Rejection
Refer to Appendix D.