3 Basic usage
9
Guide the patient's xation.
Use the external xation target to guide the patient's visual xation
until the part you want to photograph becomes visible through the op-
tical viewnder (eyepiece).
¿
Position the external xation target where the patient can x the vi-
sion with the eye not examined.
¿
If the patient's xation cannot be achieved with the eye not pho-
tographed, press the “LCD monitoring button”; then, select an
internal xation target using the “xation target selection button”,
or use the internal xation target (optional accessory) to guide the
patient's xation.
0
Adjust the focus.
Turn the “focusing knob” to adjust the focus. When you turn the focus-
ing knob, the focus dots on the LCD monitor move left and right. Turn
focusing knob so that the upper and lower focus dots come to form a
straight line.
¿
The focus dots are not displayed when using the diopter compensation lens. Adjust the focus as you
view the retinal image on the LCD monitor.
-
Start the timer.
Press the “Timer button” on the right side panel at the same time that
uorescein is intravenously injected to the patient.
Elapsed time is indicated on the LCD monitor. The elapsed time at the
time of photography is also recorded in the photographed image.
^
Insert the barrier lter.
Press the “Barrier lter button” on the right side panel and insert the
barrier lter into the light path.