Chapter 2 Configuring The Validator
60
2. If necessary, press ↓ and ↵ at the same time to unlock the
↵ key.
3. If necessary, press until VALIDATOR displays.
4. Press ↑ until EC Min displays.
5. Press + or – to scroll through the percentage values: 0% to
100%. The default is 15%.
6. Press
↵ to select the desired value.
7. Press ↓ and
↵ at the same time to lock the ↵ key, then press
PAUSE to take the printer offline.
8. Press PAUSE again to put the printer online.
Rmin
NOTE: This parameter factors into the Overall Grade (page 57).
This menu item sets the minimum reflectance value used by the
validator for pass/fail threshold used during the analysis of bar
codes. A bar code will fail if all the bar reflectance values are above
the percentage of space reflectance set by this threshold. For
example, an Rmin setting of 50% requires that at least one bar
reflectance value fall under the mid point of the reflectance value
read for the highest space reflectance.
You can set the Rmin parameter to a percentage value from 0% to
100%. The default is 50%.
1. Press to take the printer offline and place the printer in
Menu mode.
2. If necessary, press ↓ and
↵ at the same time to unlock the
↵ key.
3. If necessary, press until VALIDATOR displays.
4. Press ↑ until Rmin displays.
5. Press + or – to scroll through the percentage values: 0% to
100%. The default is 50%.
6. Press
↵ to select the desired value.
7. Press ↓ and
↵ at the same time to lock the ↵ key, then press
PAUSE to take the printer offline.
8. Press PAUSE again to put the printer online.
.
.
.
.
.
.
.
.
.