Toshiba GA-1121 Printer User Manual


 
A
A-7 Color Measurement Instruments
9. If you have not already done so, remove the shipping spacer from the measurement
page slot.
10. Insert the end of the Color Reflection Reference strip into the calibration strip
entrance on the DTP41. Center the arrow on the strip below the alignment mark (see
the diagram on page A-2), and slide the strip into the DTP41 until you feel the resis-
tance of the feed mechanism.
11. Click Read Reference in the Instrument Calibration dialog box.
The DTP41 pulls the calibration strip through automatically.
A message appears in the Status area when the calibration is finished.
12. Click Done.
13. Close the Utilities window.
This completes DTP41 calibration.
When the DTP41 requires calibration, your computer prompts you. When the
prompt message appears, click Calibrate Now and follow the instructions, starting
with step 8 on page A-6.
After calibrating the DTP41, calibrate the GA-1121 (see page 2-19).