16
Instruction manual V 6.0 - 08/2006
200 V / 50 Hz, two-phase
Nominal voltage: 200 V AC ± 10 %
Nominal frequency: 50 Hz
Power draw: 2300 VA
Automatic cutout/ 2 x 8A / 1 x 10A
Mains safety switch UL-listed
200 V / 60 Hz, two-phase
Nominal voltage: 200 V AC ± 10 %
Nominal frequency: 60 Hz
Power draw: 2300 VA
Automatic cutout/ 2 x 8A / 1 x 10A
Mains safety switch UL-listed
2.2.5 Refrigeration system - cooling chamber
Selectable temperature range: 0° C to -30° C
at ambient temperature of
22° C, with window closed
and relative humidity ≤ 60 %: -30° C ±10 % (selectable
in 1 K steps)
Control accuracy: +1 to -3 K
Electrical power draw: 1160 W
Refrigerating output at -25°C: 1370 W
Nominal pressure: 20 bar (maximum value)
Cutoff pressure: 25 bar (via Pressostat)
Safety factor: 2, as per UL 1262
Refrigerant: 1100 g, refrigerant R404A
Compressor oil: 1220 ccm alpha 22, Kyodo
Defrosting: automatic hot-gas defrosting, programmable,
2 x automatic hot-gas defrosting cycles / 24 hours.
Duration: 1 - 15 minutes, manual defrosting as
needed.
2. Characteristics of the Leica CM3600