Leica CM3600 Copier User Manual


 
32
Instruction manual V 6.0 - 08/2006
Fuse protecting the mains switch
If the mains switch fuse has blown, please call Technical
Service.
S
Fuse protecting socket panel
If the socket panel fuse has blown, please proceed as follows:
Prior to exchanging the fuse in the socket panel, switch the
instrument mains switch off and unplug the instrument from
mains. Only use fuses of the same type and characteristics as
originally supplied (see chapter 2.2 ‘Technical data’).
1. Push the fuse carrier inwards with a screw driver and turn the screw
driver approx. one quarter turn to the left.
2. Pull out the fuse carrier.
3. Remove the fuse.
4. Insert a replacement fuse of identical specification.
5 Reinsert the fuse carrier.
6. Push the fuse carrier slightly inwardswith the screw driver and turn it
one quarter turn to the right.
7. Reconnect the instrument to mains and switch back on.
5. Installation
Fuse protecting the mains
switch
Fuse protecting socket panel
Fuse carrier