Leica CM3600 Copier User Manual


 
52
Instruction manual V 6.0 - 08/2006
Dehydration date
Dehydration start time
Dehydration duration
Indication of start date and time, and duration of the automatic dehydra-
tion cycle. Clicking on one of the parameters opens the dialog box „Dehy-
dration time“.
Click on individual parameters (day, month, year, hour or minutes) to
change the corresponding settings.
The desired duration of the dehydration procedure can be selected in the
duration field. Values from 1 - 48 hours can be selected. Click OK to
confirm.
To speed up the dehydration procedure, the dehydration
cryochamber temperature can be lowered. This can be done
via the ‘Dehydration T-reduction’ indication.
Dehydration T-reduction
Indication of temperature reduction for dehydration. Click on this indica-
tion to change the currently selected value. Values from 0 - 20 K can be
selected.
Section thickness
Indication of section thickness. Click on this indication to change the
currently selected value. Values from 0 - 200 μm can be selected.
Retraction value
Indication of the currently selected retraction value. Click on this indica-
tion to change the currently selected value. Values from
50 - 250 μm can be selected.
6. Software