6.5.1 Parameter setting
Set chamber temperature
Click on this indication to change the currently selected value. Tempera-
ture values from -30 to +1° C can be selected.
Alarm temperature
Click on this indication to change the currently selected value. Tempera-
ture values from -25 to -5° C can be selected.
The alarm temperature should not be too close to the set
chamber temperature. If the alarm temperature is too close to
the set chamber temperature, an alarm can be triggered
simply when working with the instrument during a prolonged
period of time or during the daily defrost cycle.
A temperature difference of approx. 5 K (Kelvin) is recom-
mended.