Leica CM3600 Copier User Manual


 
76
Instruction manual V 6.0 - 08/2006
6.8.3 Error codes: meaning and troubleshooting
Error code Error message Trouble shooting
Error 01
Error 02
Error 03
Error 04
Error 05
Error 06
Error 07
Error 08
Error 09
Error 10
Error 11
Knife holder - limit position exceed-
ed.
Limit position of sledge exceeded
Chamber - Alarm temperature ex-
ceeded
Power failure
Data transfer failed
(happens rather frequently subse-
quent to a power failure)
Emergency stop activated
Stepper motor - step error
Refrigerating system – pressure ex-
ceeded
Flap limit switch malfunction
Call Technical Service
Re-initialize instrument. If the problem re-
curs, call Technical Service.
Re-initialize instrument. If the problem re-
curs, call Technical Service.
Check whether there has been a power
failure!
Check the selected alarm temperature val-
ue. It must differ at least 5 K from the
chamber temperature selected.
Check the cable connection between PC
and control unit.
If problem recurs, call Technical Service
Not defined
Deactivate emergency switch
Call Technical Service.
Check the air circulation between the con-
denser and the control unit.
Check the distance of the instrument to the
wall (at least 50 cm)!
If problem recurs, call Technical Service
Re-initialize instrument. If the problem re-
curs, call Technical Service
Call Technical Service.