Leica CM3600 Copier User Manual


 
38
Instruction manual V 6.0 - 08/2006
6.3 Initialization
To be able to work with the instrument, after completion of the
start and log-in procedure the microtome must be initialized.
For initialization, proceed as follows:
1. In the „Main window“, click on the INIT button blinking in yellow color
(in ‘Mode’ field).
The „Safety check - initialization“ window will open up.
2. Check wheter you can confirm each safety query - if not, carry out the
required actions. Subsequently, confirm each safety query by mouse
click on the corresponding check box.
3. Having confirmed all safety queries, click on OK.
The microtome will then be initialized:
The knife holder moves to the uppermost position.
The microtome sledge moves to the end positions on the right and left.