Appendix A: Performance Specification
A-7
1740A/1750A Series Waveform/Vector Monitor User Manual
Table A- 4: Measurement Cursors
CATEGORY
DESCRIPTION
Waveform Accuracy Req: Voltage: 0.5%
Timing: 0.5%, for line rate sweeps
RI: Typically 0.5% for field rate sweeps
Vector Accuracy Req: Gain: ±1.5%
Phase: ±1°
RI: Measured with respect to the Color
Bar signal
Table A- 5: RGB/YRGB
CATEGORY
DESCRIPTION
RGB/YRGB Req: Staircase input gain: 0.8 V/division ±10%
RI: RGB Sweep Length:
1 Field ≈30% of normal
1Line ≈ 30% of normal
RI: YRGB Sweep Length:
1 Field ≈ 25% of normal
1Line ≈ 25% of normal
RI: Maximum s taircase operating signal: DC
signal plus peakacnotto exceed --12 V
to+12 V. Line or field rate sweep.
RI: Peak-to-peak acsignal not to exceed
12 V.
RI: Sweep Repetition Rate: Field or line rate
of displayed video or external
sync signal as selected by the front-