Operator’s Checkout Procedure
1740A/1750A Series Waveform/Vector Monitor User Manual
2-21
LINE SEL
ALL131
FIELD
ALL 1OF2
1OF81OF4
1H
15H
-- -- 1 4 5
Figure 2-13. Line Select display with 15H selected, in 2 FIELD sweep.
Select 1of2. Now only the first field has intensified l ines,
and a
NEXT FIELD selection appears. Select NEXT FIELD and
note that the intensified lines now appear only in the second
field and the readout is
F2 131 F2 145.
Select MULTIPLE with WAVEFORM, VECTOR, and SCH (1750A--
Series only). Note that the displays appear in the order listed
here. Exit the line select menu. Enter the menu again and
note that the user settings are saved. Exit the menu.
10. V ectorscope Display
Select
VECTOR display mode. Adjust the right bezel control
to place the burst vector on the 180° graticule line. The
display should appear similar to Figure 2-14.
11. Picture Monitor Display
Select
PICTURE display mode. A picture monitor display of
the selected input signal appears. This can be used to visual-
ly identify the signal source. See Figure 2-15.