VXI VX1410A Computer Hardware User Manual


 
STATus Subsystem
3–44
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:OPERation:VXIMonitor:ACFail?
Queries the time when the VXI ACFAIL signal was asserted. The time is relative
to the mainframe power-on time.
If there are multiple occurrences of the ACFAIL signal, only the time of the
first assertion is saved. Reading the time enables it to record the next ACFAIL
signal assertion time.
When no ACFAIL is detected, the query returns a value of 0 in all fields. At
power-on, or restarting VXIMonitor the values of all fields are set to 0.
There is no command associated with this query.
STATus:OPERation:VXIMonitor:ACFail?
Parameter Query response
Not Applicable <Hour>,<Minute>,<Second>
0
None
None
Query: STAT:OPER:VXIM:ACF?
Response: 0,0,0
STATus:OPERation:VXIMonitor:CONDition?
STATus:OPERation:VXIMonitor:ENABle
STATus:OPERation:VXIMonitor?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands