STATus Subsystem
3–104
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:SCONdition?
Returns the pass/fail condition of all the monitors. A value of one in a bit
indicates the corresponding current condition has failed. Unused bits always
return a value of zero. The definition of the bits can be found in STATus:SEVent?
command on page 3–106.
Reading the register does not clear it.
There is no command associated with this query.
Summary information for current, power, triggers and IRQ reside in the overflow
register. If bit 30 is set, the overflow register should be queried for more
information. See the STATus:S2CONdition? query on page 3–105.
STATus:SCONdition?
Parameter Query response
Not Applicable <NR1>
0
None
None
Query: STAT:SCON?
Response: 256
STATus:SEVent?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands