VXI VX1410A Computer Hardware User Manual


 
Adjustment Procedures
7–4
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
CALibration
CALibration?
This a service-related command. The command calibrates the exhaust tempera-
ture of each slot. While the mainframe executes the command, the command sets
the CALibrating bit in the Operation Status Condition register.
The query performs the same function as the command but it also returns a pass
or fail response. A return value of 0 means that the calibration has passed and a
return value between 10 and 22 indicates a calibration failure in slots 0
through 12.
CALibration[:ALL]
CALibration[:ALL]?
<Parameter> Query response
Not Applicable <NR1>
0
10 – 22
Not Applicable
–203, “Command Protected”
You attempted to execute a command that is protected by a password. See the
SYSTem:PASSword command beginning on page 7–5 for more information.
None
Command: CAL
Query: CAL?
Response: 0
None
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands