VXI VX1410A Computer Hardware User Manual


 
STATus Subsystem
3–90
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:QUEStionable:TEMPerature:MAXimum?
Queries the maximum measurable temperatures in the mainframe since the last
factory maintenance was performed; the returned value is in °C.
There is no command associated with this query.
STATus:QUEStionable:TEMPerature:MAXimum? <Location>
<Location> Query response
{AMBient | OUT0 . . . OUT12 | DELTa0 . . . DELTa12}
AMBient – ambient input temperature
OUT0 . . . OUT12 – output temperature of each slot
DELTa0 . . . DELTa12 – output temperature minus
input temperature of each slot
<NR1>
0
–224, “Illegal Parameter Value”
You attempted to use an illegal parameter value.
None
Query: STAT:QUES:TEMP:MAX? AMB
Response: 32
STATus:QUEStionable:TEMPerature:LEVel?
STATus:QUEStionable:TEMPerature:CMAXimum?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands