VXI VX1410A Computer Hardware User Manual


 
Adjustment Procedures
7–2
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
9. Refer to Table 7–1 and set up the PC RS-232 port configurations. The
RS-232 settings must match those of the mainframe.
Table 7–1: PC RS-232 settings for the calibration
Procedure
RS-232 Parameter Setting
Baud Rate 9600
Data Bits 8
Stop Bits 1
Parity None
Flow Control Xon/Xoff
10. Make sure that the mainframe has had at least five minutes to warm up
before continuing this procedure.
11. Send the following commands to enable the password and to calibrate the
IntelliFrame Mainframe:
SYST:PASS VX1410"
CAL?
12. Check that the CAL? query returns a value of zero to indicate no error has
been detected during the calibration.
NOTE. Steps 13 and 14 are only necessary if the you have replaced the Enhanced
Monitor. If you haven’t replaced the Enhanced Monitor go to step 15.
13. Note the serial number on the rear panel of the mainframe.
14. Send the following command to set up the serial number in the Enhanced
Monitor:
SYST:SER Bxxxxxx"
The value of Bxxxxxx is the serial number as noted from the rear of the
mainframe.