SYSTem Subsystem
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–121
SYSTem:COMMunicate:SERial:SBITs
SYSTem:COMMunicate:SERial:SBITs?
Sets or queries the number of stop bits sent with each character over the RS-232
port. This command sets both the transmit and receive values.
SYSTem:COMMunicate:SERial[:RECeive]:SBITs <Stop Bits>
SYSTem:COMMunicate:SERial[:RECeive]:SBITs?
<Stop Bits> Query response
<NRf>
1
2
MINimum
MAXimum
<NR1>
1
2
1
2
Not Applicable
–222, “Data Out of Range”
You attempted to set terminator to an illegal value.
None
Command: SYST:COMM:SER:SBIT 2
Query: SYST:COMM:SER:SBIT?
Response: 2
All SYSTem:COMMunicate:SERial commands
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands