Agilent Technologies 85225F Network Card User Manual


 
84 Installation and User’s Guide
2 Installing the System
Agilent E5260A/70B Systems
To connect the bias networks
Step Action Notes
1 For Agilent E5260A/70B
systems, refer to the
following figure.
2 Connect the triaxial
cables to the E5260A/70B
SMUs.
*
a From the front of the system rack, connect
triaxial cables to the front panel high power
SMU FORCE and SENSE outputs on the
E5260A/70B.
b Connect a triaxial cable to the front panel
medium power SMU FORCE and SENSE
outputs on the E5260A/70B.
c From the front of the system, route the
cable through the upper feedthrough panel.
d From the rear of the system, route the
HPSMU cable through the port 2 hole in the
lower feedthrough panel.
e From the rear of the system, route the
MPSMU cable through the port 1 hole in
the lower feedthrough panel.
The triaxial cable model number is
16494A Option 002.
The high speed high power SMU is the
Agilent E5290A.
The high speed medium power SMU is
the Agilent E5291A.
The high power SMU is the Agilent
E5280A.
The medium power SMU is the Agilent
E5281A.
Refer to Figure 4 on page 27