Fluke RPM4 Computer Monitor User Manual


 
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATIONS
Page 117 © 2004-2007 DH Instruments, a Fluke Company
readings with PA = 0 and PM = 1. Then applying the as received PA and PM
and P
offset
values to the readings calculates as received readings (the readings
that the transducer would have made with the old PA, PM and P
offset
).
It is recommended that “as received” values of PA, PM and P
offset
(for Axxx
Q-RPTs if absolute mode AutoZero is used in normal operation) be recorded for
each Q-RPT prior to running the calibration. The current PA, PM and P
offset
(for
Axxx Q-RPTs) can be viewed by pressing [SPECIAL], <8cal>, <1view>.
5.2.2 EQUIPMENT REQUIRED
The recommended calibration standards for RPM4 Q-RPTs are DHI PG7000 piston
gauges (PG7102, PG7202, PG7601 for gas; PG7302 for oil). Contact DHI for additional
information.
5.2.2.1 GAS OPERATED Q-RPTS, A10M AND LOWER
Gas operated piston gauge (deadweight tester), with the following
characteristics:
Measurement uncertainty of ± 0.0025 % of reading for Standard Class
Q-RPTs or ± 0.002 % of reading for Premium Class Q-RPTs. A standard
with higher measurement uncertainty may be used but RPM4 measurement
uncertainty may be degraded proportionally from published specifications.
If the Q-RPT is Axxx (absolute) and will be used in the absolute and/or
negative gauge measurement modes, the reference must be able to apply
absolute pressures: Absolute pressures may be defined either by operation
relative to an evacuated bell jar or, for higher pressures, by addition of
atmospheric pressure measured by a suitable barometer. Axxx (absolute)
Q-RPTs that will not be used in absolute or negative gauge
measurement modes (as is often the case for higher pressure Q-RPTs)
do not require the application of absolute pressure for calibration and
may be calibrated using a gauge pressure standard.
Axxx (absolute) Q-RPTs calibrated in gauge measurement mode by
applying gauge reference pressure values should be used in gauge mode
only (see Sections 3.3.3, PRINCIPLE, 5.2.5).
Able to supply the recommended sequence of pressure points in the
range to be calibrated: See Section 5.2.4 for information on the
recommended calibration point sequence for various Q-RPT types
and classes.
5.2.2.2 OIL OR GAS OPERATED Q-RPTS A14M AND HIGHER
Gas or oil operated piston gauge (deadweight tester), with the following
characteristics:
Do not put oil or other liquids in a Q-RPT that is meant to be operated
with gas.