RPM4™ OPERATION AND MAINTENANCE MANUAL
© 2004-2007 DH Instruments, a Fluke Company Page 38
The Rate DISPLAY is the default display except in differential measurement
mode in which the default DISPLAY is RPT.
To go to a DISPLAY other than Rate, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).
3.3.6.3 Dev (Deviation)
PURPOSE
To activate the Deviation DISPLAY and/or set the deviation target value.
See Section 3.3.6 [DISPLAY], Principle.
OPERATION
To activate the Deviation DISPLAY press
[DISPLAY], <3Dev>. The display is:
Target:
100.0000 psi a
Edit the desired target value. Pressing [ENT] returns to the main run screen with
the Deviation DISPLAY active using the entered target value.
The target value is the value from which deviations (D) are measured by the
Deviation DISPLAY following:
D = current pressure - target pressure
With the Deviation DISPLAY active the main run screen is:
1. Deviation of current pressure from the target
value.
2. Target value.
99.1135 psi a zH3
D -0.8865 T100.0000
Pressing [ENT] from the main run screen when the Deviation DISPLAY is
active goes directly to the Target editing screen. This allows the target
value to be changed without going through the DISPLAY menu.
If the pressure measurement unit or mode is changed while the Deviation
DISPLAY is active the target value remains at the same numerical value.
It is NOT converted.
To go to a DISPLAY other than Deviation, press [DISPLAY] and make a new
DISPLAY choice (see Section 3.3.6).