PSC 8500 Barcode Reader User Manual


 
5-6 0DJHOODQ 6FDQQHU
the Field Standard Weight set on the center of the weighing area
and press the Zero Push Button again.
9. The Yellow LED will extinguish for approximately 10 seconds
and the Remote Display will alternately display and
until the scale is ready to proceed.
10. If the calibration was successful, the speaker sounds a single tone,
the Scale Status LED begins blinking again, and “End-” appears
in the Remote Display.
11. If the calibration was not successful, the speaker will sound five
tones indicating a scale failure, and the Scale Status LED will
blink twice, strobe (fast blinks) and then continually repeat this
sequence until reset. Remove all weight from the Weigh Platter,
and repeat the procedure starting with step 8.
12. Press the Zero Push Button again to permanently store the cali-
bration data and exit Calibration Mode. You have completed the
calibration of the scanner/scale. The unit’s calibration must now
be verified as required by state and/or local weights and measures
regulations. The verification procedure must be performed to
assure that a scale will pass Weights and Measures requirements
before it is placed into commercial/retail service.
You have completed the scale calibration procedure. You must now con-
tinue with the calibration verification tests to complete the scale’s calibra-
tion. If you are weighing in pounds, continue with the following
procedures to verify that the scale was successfully calibrated and that it
weighs properly. If you are weighing in kilograms, turn to the instructions
titled, Calibration Verification (Kilograms), later in this section.