TANDBERG D50104-15 Computer Monitor User Manual


 
4
Health Care System III
Warning
While no currently available technology can completely substitute for the in-person physical examination of an
individual patient or specimen, the TANDBERG products provide high quality, high resolution, long distance
images. When used properly, these Healthcare products can provide a significant and valuable tool for physicians
and other medical professionals who are unable to examine a patient or specimen in person. The use and value of
the system will vary depending on the specific circumstances of the patient’s condition. Ultimately, judgements on
how this tool should be used must be in the individual discretion of physician supervising the patient’s care.
For the customers in the USA
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own expense.
You are cautioned that any changes or modifications not expressly approved in this manual could void your
authority to operate this equipment.
Important safeguards/notices for use in
the medical environments
1) All the equipment connected to this system
shall be certified according to Standard
IEC601-1 or other IEC/ISO/CSA/UL Standards
applicable to the equipment.
2) When this system is used together with other
equipment in the patient area*, the equipment
shall be either powered by an isolation
transformer or connected via an additional
protective earth terminal to system ground
unless it is certified according to Standard
IEC601-1 and IEC601-1-1.
Note: The power outlets provided in the system
are powered from an isolation transformer.
*Patient Area
Symbols Location
Rear panel
of monitor
Rear panel
of monitor
Rear panel of
monitor and
power bar
This Symbol
Indicates
The equipotential
terminal which brings
the various parts of a
system to the same
potential.
Functional earth
terminal
Attention, consult
ACCOMPANYING
DOCUMENTS
3) The leakage current could increase when connected to other equipment.
4) The operator should take precautions to avoid touching the rear panel input and output circuitry and the
patient at the same time.
5) The main power switch of the system is not easily accessible. To isolate the system from the mains supply
remove the mains plug from the wall socket.