Xantrex Technology XDC 20-600 Power Supply User Manual


 
SCPI Commands for Digital Interfaces
TM-XDOP-01XN 4–23
Last Setting
Selecting Last Setting lets you restore the settings that are in use when the power
supply is powered off, the next time it is powered on.
To power on the last stored setting:
OUTP:PON:REC LAST
To check the current user setting:
OUTP:PON:REC?
The response should be LAST.
User Setting
Selecting User Setting lets you restore a custom setting next time the unit is
powered on. This assumes that at least one user setting has been set up and stored
in memory. See “Store User Settings” on page 3–21.
To power on user setting #1:
OUTP:PON:REC USER1
To check the setting:
OUTP:PON:REC?
The response should be USER1.
The user setting must be in the range of 1–10. See “Store User Settings” on
page 4–25 for an explanation of how to save user settings.
Auto Sequence
Selecting Auto Sequence lets you recall a stored program next time the unit is
powered ON. This assumes that at least one program has been created and stored
in memory. See “Editing Trigger Source of a Sequence” on page 3–41 for more
information. See for information about remotely programming and running auto
sequence programs.
To reset to user setting #1:
OUTP:PON:REC SEQ1
To check the setting:
OUTP:PON:REC?
The response should be SEQ1.