Xantrex Technology XDC 20-600 Power Supply User Manual


 
Power Supply Operation
TM-XDOP-01XN 3–13
To set the Over-Power Protection:
To set the Under-Power Protection:
Step # Do This You Will See
1 OVP SET 0V
2
Press repeatedly until the desired setting
appears.
OPP SET 0W
3 OPP SET 0W
4 Use both the Current and Voltage knobs, or the
numeric keypad, or the arrow keys to enter a value.
The value must be between 0W and 103% of the
unit’s rated power.
OPP SET ####W
5 S/D if trip? N
6 Use the arrow keys to select Yes or No. For this
example, Yes is selected.
S/D if trip? Y
7 This setting is saved and
the display returns to its
default operating mode.
PROT SET
ALARMS
PROT SET
ALARMS
ENTER
ENTER
ENTER
Step # Do This You Will See
1 OVP SET 0V
2
Press repeatedly until the desired setting
appears.
UPP SET 0W
3 UPP SET 0W
4 Use both the Current and Voltage knobs, or the
numeric keypad, or the arrow keys to enter a value
(must be between 0W and 103% of the unit’s rated
power).
UPP SET ####W
5 S/D if trip? N
6 Use the arrow keys to select Yes or No. For this
example, Yes is selected.
S/D if trip? Y
7 This setting is saved and
the display returns to its
default operating mode.
PROT SET
ALARMS
PROT SET
ALARMS
ENTER
ENTER
ENTER