Xantrex Technology XDC 20-600 Power Supply User Manual


 
SCPI Command Reference
C–32 TM-XDOP-01XN
Table C-15
Legacy Commands
Function SCPI Command Description Query
Reset
CLR
Equivalent to *RST and SYSTem:RESet commands
Performs a device reset. Set the power supply to a known state that is
independent of the use history of the device
N/A
Query System Error
ERR?
Equivalent to SYSTem:ERRor? command except that the return string
contains the command
Returns the next error in the instrument’s error queue
N/A
Identification Query
ID?
Query identification string. (Model ID and Version) N/A
Read Output
Current
IOUT?
Equivalent to MEASure:CURRent? command
Read output current
N/A
Set High Current
Limit
IMAX <current>
Equivalent to SOURce:CURRent:LIMit:HIGH <current> command.
Set upper limit of current setpoint range (soft limits)
Yes
Set (Immediate)
Current Setpoint
ISET <current>
Equivalent to SOURce:CURRent command.
Change current setpoint
Yes
Set Output State
OUT <on-off-
state>
Equivalent to OUTP <on-of-state> command.
Enable/disable the power supply output
Yes
Set Over Voltage
Protection Level
OVSET <voltage>
Equivalent to SOURce:VOLTage:PROTection:OVER:LEVel <voltage>
command.
Set the over voltage protection level
Yes
ROM Query
ROM?
Queries the main firmware version N/A
Clear Output
Protection
RST
Equivalent to OUTPut:PROTection:CLEar command.
Clears the protection mechanism
N/A
Set High Voltage
Limit
VMAX <voltage>
Equivalent to SOURce:VOLTage:LIMit:HIGH <voltage> command.
Set upper limit of voltage setpoint range (soft limits)
Yes
Read Output
Voltage
VOUT?
Equivalent to MEASure:VOLTage? command.
Read output voltage
N/A
Set (Immediate)
Voltage Setpoint
VSET <voltage>
Equivalent to SOURce:VOLTage command.
Change voltage setpoint
Yes