Xantrex Technology XDC 20-600 Power Supply User Manual


 
Calibration (6000 Watt only)
B–10 TM-XDOP-01XN
7. Power supply calculates and stores calibration constants.
8. The menu will go to the 5V analog current readback calibration menu.
Analog Programming Interface Current Readback Calibration
1. Attach a DVM across the current readback lines, Pins B6 and B1(GND).
2. Enter 5V analog current readback calibration menu
Select ANLG I RB 5V from the calibration menu. Press ENTER.
You will be prompted to set up for analog current readback calibration. Press
ENTER when ready.
3. Minimum calibration level
The power supply will automatically set the current readback lines to
approximately 10% of full scale.
4. Enter voltage data
Enter the voltage read from the external DVM. Press ENTER
5. Maximum calibration level
The power supply will automatically set the current readback lines to
approximately 90% of full scale.
6. Enter voltage data
Enter the voltage at the current readback lines, read from the external DVM.
Press ENTER.
7. Power supply calculates and stores calibration constants.
8. The menu will go to the 10V analog voltage programming calibration menu.
Analog Programming Interface 0-10V Range
The 0-10V range of the analog programming interface must be calibrated
separately. Follow the procedure exactly as for calibrating the 0-5V range, using
the corresponding 10V menu options. All the analog signals will be scaled by a
factor of 2.
TM-XDOP-01XN.book Page 10 Monday, July 17, 2006 11:19 AM