Fluke 2640A Network Card User Manual


 
2640A/2645A NetDAQ
Users Manual
6-12
3. Verify Accuracy Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).
Volts DC Range 5700A Output Minimum Reading Maximum Reading
90 mV Short Circuit (Zero) -0.000008V +0.000008V
90 mV +90 mV +0.089980V +0.090020V
90 mV -90 mV -0.090020V -0.089980V
300 mV Short Circuit (Zero) -0.000017V +0.000017V
300 mV +300 mV +0.299944V +0.300056V
300 mV -300 mV -0.300056V -0.299944V
3V Short Circuit (Zero) -0.00015V +0.00015V
3V +3V +2.99946V +3.00054V
3V -3V -3.00054V -2.99946V
30V Short Circuit (Zero) -0.0017V +0.0017V
30V +30V +29.9944V +30.0056V
30V -30V -30.0056V -29.9944V
300V Short Circuit (Zero) -0.017V +0.017V
300V +300V +299.944V +300.056V
300V -300V -300.056V -299.944V
4. Close Spy Window.
Volts DC Accuracy Test (2645A) 6-10.
Complete the following procedure to test the accuracy of the volts dc function for
the 2645A. Measurement accuracy applies to all channels, not just the channel
used for the test.
1. Configure Channel 1 for Volts DC In NetDAQ Logger, configure channel 1
for volts dc, 90 mV range. (See "Configuring Analog Channel Functions" in
Chapter 3 of this manual.)
2. Open Spy Window Select the Spy command from the Utilities menu. Select
channel 0101 (instrument 01, channel 01) from the Channel list. Click OK to
open the Spy window.