Maintenance
Performance Test
6
6-19
Resistance Range* Decade Resistor Minimum Reading Maximum Reading
30 k
Short Circuit (Zero)
700 1 k
30 k 29 k 29.681 k 30.019 k
300 k 290 k 289.07 k 292.63 k
3 M 2.9 M 2.8607 M 2.9410 M
* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance plus 0.01% decade resistance tolerance. You must add any additional lead wire
resistance present in your setup to the resistance values given in this table.
Resistance Range* 5700A Minimum Reading Maximum Reading
30 k
Short Circuit (Zero)
700 1 k
30 k 19 k 19.686 k 20.014 k
300 k 190 k 189.60 k 192.10 k
3 M 1.9 M 1.8740 M 1.9277 M
* The resistance accuracy in this table makes allowance for up to 0.1 Ohm of lead wire
resistance. You must add any additional lead wire resistance present in your setup to the
resistance values given in this table.
5. Close Spy Window.