Fluke 2640A Network Card User Manual


 
Specifications
2645A Specifications
A
A-21
2645A 4-Wire Resistance Measurement Specifications A-24.
Tables A-30 to A-32 provide 2645A specifications for the 4-wire resistance
measurement function. The 4-wire measurements use 2 input channels a decade
apart, e.g., channels 4 and 14.
Table A-30. 2645A 4-Wire Resistance Temperature Coefficient
Specification Characteristic
Temperature Coefficient Add 1/10th the 90 day specification per ºC
above 28ºC or below 18ºC.
Accuracy at -20ºC Multiply the -10°C to + 60°C accuracy
specification by 2. After 1 hour warm-up. For
accuracy between -10°C and -20°C, interpolate
linearly.
Table A-31. 2645A 4-Wire Resistance Range and Resolution Specifications
Range
Resolution
Current
Applied
Full Scale
Voltage
Maximum
Voltage
Slow Fast
Applied
300 10 m 30 m 1 mA 300 mV 3.5V
3 k 100 m 300 m 100 µA 300 mV 3.5V
30 k 1 3 10 µA 300 mV 3.5V
300 k 10 30 10 µA 3.0V 3.5V
3 M 100 300 1 µA 3.0V 3.5V
Table A-32. 2645A 4-Wire Resistance Accuracy Specifications
Accuracy, 3σ ± (% input + V)
18ºC to 28ºC-10ºC to 60ºC
Range
90 Day 1 Year 1 Year
Slow Fast Slow Fast Slow Fast
300 .02%+60 m .02%+.1 .02%+.1 .02%+.2 .084%+.25 .084%+.42
3 k .02%+.6 .02%+2 .02%+1 .02%+3 .084%+2.5 .084%+8.4
30 k .02%+6 .2%+200 .02%+10 .2%+300 .084%+25 .84%+840
300 k .5%+80 1%+2 k .5%+150 1%+3 k 2.1%+336 4.2%+8.4 k
3 M 1.3%+1 k 2%+120 k 1.3%+2 k 2%+200 k 5.46%+4.2 k 8.4%+200 k