Fluke 2640A Network Card User Manual


 
2640A/2645A NetDAQ
Users Manual
A-12
2640A 4-Wire Resistance Measurement Specifications A-15.
Tables A-15 to A-17 provide 2640A specifications for the 4-wire resistance
measurement function. The 4-wire measurements use 2 input channels a decade
apart, e.g., channels 4 and 14.
Table A-15. 2640A 4-Wire Resistance Temperature Coefficient
Specification Characteristic
Temperature Coefficient Add 1/10th the 90 day specification per ºC
above 28ºC or below 18ºC.
Accuracy at -20°C Multiply the -10°C to + 60°C accuracy
specification by 2. After 1 hour warmup. For
accuracy between -10°C and -20°C, interpolate
linearly.
Table A-16. 2640A 4-Wire Resistance Range and Resolution Specifications
Resolution
Current Full Scale Maximum voltage
Range
Slow Fast
Applied Voltage applied by instrument
300 1 m 3 m 1 mA 300 mV 3.5V
3 k 10 m 30 m 100 µA 300 mV 3.5V
30 k 100 m 300 m 10 µA 300 mV 3.5V
300 k 1 3 10 µA 3.0V 3.5V
3 M 10 30 1 µA 3.0V 3.5V
Table A-17. 2640A 4-Wire Resistance Accuracy Specifications
Accuracy, 3σ + (% input + V)
18ºC to 28ºC-10ºC to 60ºC
Range
90 Day 1 Year 1 Year
Slow Fast Slow Fast Slow Fast
300 .015%+20 m .02%+80 m .02%+50 m .02%+120 m .084%+126 m .084%+336 m
3k .02%+.3 .02%+.8 .02%+.5 .02%+1.2 .084%+1.26 .084%+3.36
30 k .03%+3 .04%+10 .03%+5 .04%+15 .126%+12.6 .168%+42
300 k .1%+40 . 2%+100 .1%+60 .2%+150 .42%+168 .84%+420
3 M* .25%+800* .5%+10k* .25%+1 k*.5%+1.5k* 1.05%+3.36 2.1%+4.2 k*
* The 3 M range is susceptible to the absorption of humidity under extreme conditions. If the instrument is
operated normally within its specified temperature-humidity range, the 3 M range meets its accuracy
specifications. However, if the instrument is “soaked” at 50°C, 90% RH the 3M range may require 1 hour
of “dry-out” time at 25°C, <40% RH for each hour of “soak” time in order to achieve its specified accuracy.