Kodak DryView 8900 Printer User Manual


 
1-16 5E6155 January 10, 2006
Safety and Related Information
Note 1 Field strengths from fixed transmitters, such as base station for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Kodak DryView 8900 Laser Imager is used exceeds the applicable RF
compliance level above, the Kodak DryView 8900 Laser Imager should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Kodak DryView 8900 Laser Imager.
Note 2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 v/m.
Recommended separation distance between portable and mobile RF communications
equipment and the Kodak DryView 8900 Laser Imager
The Kodak DryView 8900 Laser Imager is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Kodak DryView 8900 Laser Imager can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication
equipment (transmitters) and the Kodak DryView 8900 Laser Imager as recommended below, according to the
maximum output of the communications equipment.
Rated maximum output
power of transmitter (P)
in Watts (W)
Separation distance (d) according to frequency of transmitter
in meters (m)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01
0.1
1
10
100
Guidance and Manufacturer’s Declaration for Electromagnetic Immunity
d 1.17 P=
d 1.17 P=
d 2.33 P=